The
soxhlet distillation extraction method is used to dissolve and extract oil and
brine from rock core samples
by using solvents. The cleanliness of the sample is determined from the color of
the solvent that siphons periodically from the extractor which must be clear.The samples are placed in the extractor and cleaned by refluxing solvent.
The solvent is heated and vaporized in boiling flasks and cooled at the top by
condenser. The cooled solvent liquid falls into the sample chamber. The cleaned
solvent fills the chamber and soaks the core sample. When the chamber is full,
the dirty solvent which was used to clean the core siphons back into the boiling
flask and is redistilled againThe apparatus consists of a distillation/extraction glassware unit and a
heating mantle with thermostatic controller. The glassware is composed of
boiling flask, soxhlet extractor and condenser. A flexible plastic tubing is
also used to connect the condenser to the water cooling unit. All these devices
are mounted on a mounting rack.
Features:
Soxhlet
for Plug Sized Core Samples
Configuration
single unit model/multiple
units model (up to 6 extractors)
Core Sample Diameter
up to 1.5 inches
Core Sample Length
up to 4 inches
Extractor Capacity
250 ml
Boiling Flask
500 ml
Heating Temperature
up to 450 degrees C
Water Cooling System
0.5 to 3 liters per minute, 18
degrees C
Power Supply
220 VAC 50 Hz
Soxhlet
for Full Sized Core Samples
Configuration
single unit model/multiple
units model (up to 6 extractors)
Core Sample Diameter
up to 4 inches
Core Sample Length
up to 6 inches
Extractor Capacity
3 liters
Boiling Flask
6 liters
Heating Temperature
up to 450 degrees C
Water Cooling System
0.5 to 3 liters per minute, 18
degrees C
Power Supply
220 VAC 50 Hz
VINDUM
ENGINEERING, INC. 1 Woodview Court
San Ramon, California 94582-2307 U.S.A.
Telephone:
+1.925.275.0633
Facsimile: +1.925.275.9697 info@vindum.com
VINDUM
ENGINEERING, INC. Canada
Telephone: +1.403.770.0758 or 281.782.8312
Facsimile: +1.630.845.8437 christa@vindum.com