The
Dean Stark extractor is used to determine fluid saturation of core samples. The
sample is first weighed and then introduced into the Dean Stark extractor. The
water fraction is vaporized by boiling solvent, condensed and then collected in
a calibrated receiver. Vaporized solvent also condenses, soaks the sample, and
extracts the oil. The sample is then oven dried and weighed. The oil content is
determined by gravimetric difference while the water content is volumetrically
measured by the receiving tube. The apparatus consists of a
distillation/extraction glassware unit and a heating mantle with thermostatic
controller. The glassware is composed of boiling flask with extractor, a sample
support screen, volumetrically graduated water receiving tube, condenser and
desiccant drying tube. A flexible plastic tubing is also used to connect the
condenser to the water cooling unit. All these devices are mounted on a mounting
rack
Features:
Dean Stark
for Plug Core Sample
Configuration
Single unit model / multiple
units model (up to 6 extractors)
Core sample diameter
up to 1.5 inches
Core sample length
up to 4 inches
Extractor capacity
250 ml
Boiling flask
500 ml
Receiving tube
5 and 10 ml, 0.1 ml graduation
Heating temperature:
up to 450 degrees C
Water cooling system
0.5 to 3 liters per minute, 18
degrees C
Power supply
220 VAC 50 Hz
Dean Stark
for Full Size Core Sample
Configuration
Single unit model / multiple
units model (up to 6 extractors)
Core sample diameter
up to 4 inches
Core sample length
up to 6 inches
Extractor capacity
3 liters
Boiling flask
6 liters
Receiving tube
40 ml, 0.2 ml graduation with
drain valve
Heating temperature:
up to 450 degrees C
Water cooling system
0.5 to 3 liters per minute, 18
degrees C
Power supply
220 VAC 50 Hz