PVT Oil
Research, Core Analysis and Supercritical Fluids
Operating
Principle
Our industry leading viscosity technology is based on a
simple and reliable electromagnetic concept. Two coils move the
piston back and forth magnetically at a constant force.
Proprietary circuitry analyzes the piston's two-way travel time to
measure absolute viscosity. Built-in temperature detector (RTD)
senses the actual temperature in the sampling chamber.
Fluid flows through 1/4 inch OD, 0.086 inch tube for
measurement. Fluid may be measured under low flow or static
conditions. System measures viscosity, temperature and temperature
compensated viscosity. It provides data via RS-232, RS-485 or 4-20 mA
signal.
Equipment
Description
The model SPL440 installs into a high-pressure pump
through type system. Both ends mate to HiP HF4 fittings with a
9/16-28 thread. The
SPL440 is rated for continuous operation up to 20,000 psi and is static
tested to 30,000 psi.